کوره با تکنولوژی ایمنی مطابق با EN 1539 مدل TR 60 - TR 1050 ساخت کمپانی Nabertherm
With their maximum working temperature of up to 300 °C and forced air circulation, the ovens achieve a perfect temperature uniformity which is much better than in ovens of most competitors. They can be used for various applications such as e.g. drying, sterilizing or warm storing. Ample warehousing of standard models provides for short delivery times.
Tmax 300 °C
Working temperature range: + 5 °C above room temperature up to 300 °C
Ovens TR 60 - TR 240 designed as tabletop models
Ovens TR 450 and TR 1050 designed as floor standing models
Horizontal, forced air circulation results in temperature uniformity better than +/- 5 °C
Stainless steel chamber, alloy 304 (AISI)/(DIN material no. 1.4301), rust-resistant and easy to clean
Large handle to open and close the door
Charging in multiple layers possible using removeable grids (number of removeable grids included, see table to the right)
Large, wide-opening swing door, hinged on the right with quick release for models TR 60 - TR 450
Double swing door with quick release for TR 1050
TR 1050 equipped transport rollers
Infinitely adjustable exhaust at the rear wall with operation from the front
PID microprocessor control with self-diagnosis system
Solid state relays provide for lownoise operation
Defined application within the constraints of the operating instructions
Additional equipment
Over-temperature limiter with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the oven and load
Infinitely adjustable fan speed of the air circulation fan
Window for charge observing
Further removeable grids with rails
Side inlet
Stainless steel collecting pan to protect the furnace chamber
Safety technology according to EN 1539 for charges containing liquid solvents (TRS) up to model TRS 240, achievable temperature uniformity +/- 8 °C
Transport costors for model TR 450
Various modifications available for individual needs
Upgrading available to meet the quality requirements of AMS 2750 E or FDA
Process control and documentation with Controltherm MV software package
Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Connected | Electrical | Weight | Grids in- | Grids | Max. | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
°C | w | d | h | in l | W | D | H | load kW2 | connection* | in kg | cluded | max. | total load1 | |
TR 60 | 300 | 450 | 380 | 350 | 60 | 700 | 650 | 690 | 3.1 | 1-phase | 90 | 1 | 4 | 120 |
TRS 60 | 260 | 450 | 360 | 350 | 57 | 700 | 680 | 690 | 6.3 | 3-phase | 92 | 1 | 4 | 120 |
TR 120 | 300 | 650 | 380 | 500 | 120 | 900 | 650 | 840 | 3.1 | 1-phase | 120 | 2 | 7 | 150 |
TRS 120 | 260 | 650 | 360 | 500 | 117 | 900 | 680 | 840 | 6.3 | 3-phase | 122 | 2 | 7 | 150 |
TR 240 | 300 | 750 | 550 | 600 | 240 | 1000 | 820 | 940 | 3.1 | 1-phase | 165 | 2 | 8 | 150 |
TRS 240 | 260 | 750 | 530 | 600 | 235 | 1000 | 850 | 940 | 6.3 | 3-phase | 167 | 2 | 8 | 150 |
TR 450 | 300 | 750 | 550 | 1100 | 450 | 1000 | 820 | 1440 | 6.3 | 3-phase | 235 | 3 | 15 | 180 |
TR 1050 | 300 | 1200 | 630 | 1400 | 1050 | 1470 | 955 | 1920 | 9.3 | 3-phase | 450 | 4 | 14 | 250 |
1Max load per layer 30 kg *Please see page 60 for more information about supply voltage 2If EN 1539 is ordered connected load will increase |